Advanced Material Technology (AMT) / HALO Elevate C18

HALO Elevate

The HALO Elevate series comprises HPLC columns from Advanced Materials Technology (AMT) featuring an organosilane-modified surface. This hybrid material is therefore stable at pH values between 2 and 12. Thanks to this high pH stability, bases can be analysed in their uncharged state, resulting in good peak symmetry. The modified surface is based on AMT’s well-known Fused-Core technology, enabling high flow rates to be achieved in combination with high pH stability. Thanks to the choice of different stationary phases, such as C18, C8 and Phenyl-Hexyl, a wide range of different applications can be run at high pH values.

Products

Sort By sort-descending

Items 61-70 of 146

Page
per page
AMT-92273-108
€565.00
AMT-92272-108
€565.00
AMT-9227A-708
€8,550.00
AMT-92270-708
€2,850.00
AMT-92274-908
€1,115.00
AMT-92274-708
€850.00
AMT-92274-608
€760.00
AMT-92274-408
€680.00
AMT-92274-308
€585.00
AMT-92273-908
€1,115.00
Sort By sort-descending

Items 61-70 of 146

Page
per page

Technical Data

Phase

Modification

Particle size

Pore size

Surface area

Carbon content

End-capped

USP

pH range

Tmax

HALO Elevate C18

C18

2 µm

2.7 µm

5 µm

120 Ã…

70 m²/g

75 m²/g

60 m²/g

5.2 %C

5.6 %C

4.5 %C

+

L1

2.0–12.0

60°C

HALO Elevate C8

C8

2 µm

2.7 µm

5 µm

120 Ã…

70 m²/g

75 m²/g

60 m²/g

3.8 %C

4.2%C

3.2 %C

+

L7

2.0–12.0

60°C

HALO Elevate Phenyl-Hexyl

Phenyl-Hexyl

2 µm

2.7 µm

5 µm

120 Ã…

70 m²/g

75 m²/g

60 m²/g

4.5 %C

5.1 %C

4.0 %C

+

L11

2.0–12.0

60°C

a: for low and high pH ranges

What are the benefits of the HALO Elevate range?

Due to the organo-silane modified surface, these columns are very stable at high pH values. Amitriptyline was analysed at a pH value of 10. It was found that the retention only shifted by 1% after 500 injections (or 20,000 column volumes).

 

Test conditions

Column: HALO 120Å Elevate, 2.7 µm, 50 × 2.1 mm

Analytes: 1. Uracil 2. Acenaphthene 3. Amitriptyline

Mobile phase A: 95/5 10 mM ammonium bicarbonate pH 10 / acetonitrile

Mobile phase B: Acetonitrile

Gradient: 

Time / min

%B

0.0

0

2.5

95

3.0

95

3.1

0

5.0

0

Flow rate: 0.8 mL/min

Back pressure: 220 bar

Temperature: 60°C

Detection: UV/PDA, 260 nm

Injection volume: 1.0 µL

LC system: Shimadzu Nexera X2

Changes in selectivity due to pH

Thanks to the wide usable pH range of the HALO Elevate series, the selectivity of acidic, basic and neutral compounds can be controlled via the pH value.

  • Low pH: Bases are charged and elute rapidly
  • Neutral pH: Bases and acids are charged and exhibit rapid elution
  • Basic pH: Bases are unloaded and exhibit late elution

HALO Elevate C18

HALO Elevate C8

HALO Elevate Phenyl-Hexyl

Applications

When would you benefit from the HALO Elevate range?

  • Whenever a high pH value is advantageous (alkaline analytes)

Analysis of Omeprazole

When analysing omeprazole (pka 9.3), related compounds and impurities, the run time can be reduced by more than 84% compared to the original USP method. All 9 component are baseline separated.

 

AMT HALO Elevate C18 Omeprazole Application

Peak identities

1. Related Compound F & G 2. Related Compound B 3. Related Compound E 4. Related Compound A 5. Impurity B 6. Omeprazole 7. Impurity H 8. N'-Methyl Omeprazole 9. Impurity C

Test conditions

Column: HALO 120Å Elevate C18, 2.7µm, 150x2.1mm (92272-702)

Analysis of Tricyclic Antidepressants

Tricyclic antidepressants (TCAs) are basic drugs that can be separated with the HALO Elevate C18 in less than 3 min with excellent peak symmetries.

 

AMT HALO Elevate C18 Antidepressiva Application

Peak Identities

 

1. Nortriptyline  2. Doxepin  3. Imipramine  4. Amitriptyline  5. Trimipramine

Conclusion: With the HALO Elevate series, which is based on hybrid technology, basic analytes can be analysed in their uncharged state. This prevents secondary interactions, thereby ensuring high peak symmetry. This column is ideal for laboratories with a high volume of basic applications.

Downloads

Contact Us

Fields marked with * must be filled in

The right column for you - we will be happy to support you individually

Competent consultants are always at your side. Write a message to our consultants, we will get back to you and give you individual support.

You will find:

Application support Application support
Availability and delivery times Availability and delivery times
Price information with individual conditions Price information with individual conditions
Certificates of analysis, MSDS, data sheets Certificates of analysis, MSDS, data sheets