Troubleshooting

Tailing peaks, altered retention times, pressure fluctuations or missing detector signal - these and many more are all cases that are undesirable in the everyday life of the analyst, but nevertheless occur from time to time.

 

In order to help you with troubleshooting, some manufacturers have produced very helpful brochures on the subject of troubleshooting, which give you a good overview of possible sources of error and how to rectify them in everyday HPLC work. Below you will find possible causes and solutions for the most common problems.

We are happy to answer any questions and provide support on the subject of troubleshooting.

Technical Data

Peak tailing

Peak tailing is one of the most common problems in HPLC. The problem can have various causes. It is important to distinguish whether the problem affects all peaks or only individual peaks. If all peaks are affected, this indicates a physical problem in the system or column. This could be dead volume, a blocked frit or similar. If only individual peaks are affected, this indicates chemical problems, such as the tailing of bases on older columns.

 

ObservationPossible causeSolution/Prevention
Peak tailing with basesUndesirable interaction of bases with active silanols

1. use a more modern phase to reduce active silanols

2. add basic additives to the mobile phase to mask silanols (not necessary with modern phases)

Peak tailing of analytes that can form chelate complexes with metalsUnwanted interaction of analytes with metals in the stationary phase1. use a modern phase to reduce metals
Peak tailing with ionic/ionisable analytesUnsuitable pH value in the mobile phase leads to undesirable interactions with charged silanols

1. Check pka values of analytes and select pH value so that the analyte is fully charged or neutral

2. Use a pH value of 2.5 so that the silanols are uncharged to prevent ionic interactions

Peak tailing for all peaksBlocked frit

1. reverse column rinse (please observe manufacturer's information beforehand)

2. use inline filter

Peak tailing at all peaksVoid formation in the column

Reverse rinse 1st column (please observe manufacturer's information beforehand)

Replace 2nd column

Peak tailing at all peaksDead volume in the system

1. check all connections (fittings, correct cutting of capillaries)

2. reduce the number of connections

3. use shorter capillaries


Peak Fronting

In addition to peak tailing, fronting is the second deviation from the ideal Gaussian shape of a peak. However, fronting does not occur as frequently as tailing. The most common reasons for fronting are injection volumes that are too large and an unsuitable solvent for the injection.

 

ObservationPossible causeSolution/Prevention
Peak fronting for all peaksInjection volume or concentration too high, meaning that not every analyte can interact with the stationary phase and is eluted more quickly

1. reduce injection volume

2. reduce the concentration of the analyte

Peak fronting for all peaksInjection solvent is not compatible with the mobile phase (too strong or immiscible)

1. use mobile phase as injection solvent

2. use injection solvent that is weaker than the mobile phase

Peak fronting for all peaksChannel formation in the column due to improper handling (pH too high, pressure too high)

1. observe the recommended pH and pressure range of the column

2. exchange column and replace with a column that is compatible with the desired pH and pressure range


Peak splitting

Peak splitting can look different. If the peaks do not split very strongly, the peak splitting can appear as a broad peak. A shoulder of peaks is also possible and if the problem is very pronounced, two peaks may even be visible. The reason for peak splitting is usually a dead volume in the system or a partially blocked frit in the column and/or precolumn.

 

ObservationPossible causeSolution/Prevention
Peak splitting for all peaksPartially blocked frit in the column

1. reverse column rinse (please observe manufacturer's information beforehand)

2. use inline filter

3. if possible: replace the inlet frit of the column (please observe the manufacturer's information beforehand)

Peak splitting for all peaksPartially blocked frit in the pre-column

1. replace pre-column

2. use inline filter

Peak splitting for all peaksChannel formation in the column due to improper handling (pH too high, pressure too high)

1. observe the recommended pH and pressure range of the column

2. exchange column and replace with a column that is compatible with the desired pH and pressure range

Peak splitting stronger with early eluting peaksInjection solvent is not compatible with the mobile phase (too strong or immiscible)

1. use mobile phase as injection solvent

2. use injection solvent that is weaker than the mobile phase

Peak splitting with a peakCoelution of an impurity1. improve sample preparation
Peak splitting at one peakCoelution of a second peak1. optimise method to improve resolution
Peak splitting with ionisable compounds (acid/base)pH value too close to the pka value of the analytes or mobile phase not sufficiently buffered

1. select pH value 2 units away from pka value

2. increase buffer concentration


Additional peaks

The problem with additional or unexpected peaks can have various causes. These peaks are often late eluting analytes from a previous sample or so-called memory effects from the injection process. However, so-called ghost peaks can also be the cause of additional peaks or impurities in the sample. So-called spikes are another problem. These are not "broad" peaks but short spikes.

 

ObservationPossible causeSolution/Prevention
One or more additional peaksImpurities in the sample1. optimise sample preparation process
One or more additional peaksImpurities in the mobile phase

1. use HPLC grade solvent

2. use Ghost Guard columns

One or more additional peaksContamination in the column/pre-column

1. rinse column

2. replace the precolumn

3. in case of frequent occurrence: Optimise method/introduce rinsing step

One additional peakLate eluting peak from previous sample

1. optimise method for the sample

2. rinse column after the run

Additional peaks comparable with other/previous sampleMemory effect from the injection process

1. rinse injection system (sample loop, syringe)

2. optimise method parameters for the injection

SpikesAir in the mobile phase or in the system

1. degas mobile phases

2. check screw connections for leaks

3. if necessary, install a counter-pressure capillary on the detector (please note the maximum pressure of the detector cell)

SpikesAir in the column (column not sealed with end plug)

1. always store column with stopper

2. flush the column

 

 


Negative peaks

Negative peaks are almost always due to the detector. The mobile phase shows a higher signal than the analyte, resulting in negative peaks during detection.

 

ObservationPossible causeSolution/Prevention
Negative peaks with the RI detectorRefractive index of analyte lower than that of the mobile phase1. use mobile phase with lower refractive index
Negative peaks in the UV detectorUV adsorption of the analytes is lower than that of the mobile phase

1. change detection wavelength

2. use mobile phase with lower UV adsorption

All peaks negativeSignal polarity incorrectly set1. check polarity of the detector
All peaks negativeDetector cable connected incorrectly1. check wiring

Missing peaks

Missing peaks can be another problem with peaks. The causes can be quite simple, such as the wrong solvent or no injection, but even a very strong tailing can make it appear as if no peak is eluting. Coelution is also conceivable if a peak is sensitive to a parameter such as pH or temperature.

 

ObservationPossible causeSolution/Prevention
No peak elutedIncorrect solvent/mobile phase1. check solvent and mobile phase
No peak elutedNo injection has taken place1. check injection system (tightness)
No peak elutedWrong/no sample injected

1. check whether the vial is full enough

2. check whether the correct vial was used

No peak elutedWrong column selected1. check the column used
No peak elutedIncorrect flow rate set on the pump1. check the flow rate
No peak elutedLeak in the system, causing the flow rate to decrease1. check connections
No peak elutedIncorrect sample preparation1. check whether sample preparation was carried out correctly
No peak elutedDetector set incorrectly1. check detector settings such as wavelength (UV) or mode (negative, positive for MS)
One peak does not eluteMatrix effects interfere with the sample

1. optimise sample preparation

2. optimise separation method

One or more peaks do not eluteAdsorption in the vial (interaction with glass)1. use polymer vial or silanised vial
One or more peaks do not eluteRetention of analytes too strong

1. optimise method

2. rinse column with strong solvent


Retention time shift

If there is a shift in the retention time, it is important to know whether this shift occurs gradually or whether it occurred from one run to the next. This allows conclusions to be drawn about the cause. A constant shift, for example, indicates an ageing process of the column. If the shift occurs suddenly, this is more likely to be due to a problem in the system.

 

ObservationPossible causeSolution/Prevention
Retention time gradually decreasesLoss of the bound phase due to low pH value

1. observe the recommended pH range of the column

2. exchange column or replace with column with different pH stability

Retention time gradually increases for polar/ionic compoundsLoss of the bound phase or endcapping allows increased interaction with silanols, often accompanied by increasing tailing

1. observe the recommended pH range of the column

2. exchange column or replace with column with different pH stability

Retention time varies when using a new columnDifferences in different lots of the stationary phase (can sometimes be quite large)

1. check whether a column from a different lot was used

2. if deviations are too large, use original lot or adjust method

Retention has suddenly increased or decreasedChanges in the flow rate

1. check whether the correct flow rate is set

2. check for leaks in the system

3. check for air bubbles in the system

Retention has suddenly decreased or increasedColumn with a different dimension inserted1. check inserted column
Retention has suddenly decreased or increasedColumn was overloaded and this leads to a disturbed peak shape, whereby the retention time also deviates

1. check injection volume and concentration

2. check the column used (internal diameter)

Retention has suddenly decreased or increasedIncorrect composition of the mobile phase

1. check the composition of the mobile phase

2. if a gradient valve is used, check all valves for functionality

Retention of an ionisable compound fluctuates greatlypH value too close to the pka value of the analytes or mobile phase not sufficiently buffered, so that the smallest differences in pH value lead to large changes in retention

1. select pH value 2 units away from the pka value

2. increase buffer concentration

Retention time varies from day to day or from morning to eveningColumn temperature varies due to not tempering the column

1. use column oven

2. operate column oven above room temperature to ensure stable retention in non-air-conditioned laboratories

Retention time of the new column differs from the old columnColumn is not sufficiently equilibrated; this is particularly the case if TFA is used in the mobile phase

1. equilibrate the column for longer (in the case of TFA also overnight)

2. if the storage time is short, store the column in the mobile phase with TFA


Pressure too high

Excessive back pressure is a very common problem in HPLC and can have a variety of causes. It is important to differentiate whether excessive pressure occurs slowly or suddenly. A slow increase in pressure indicates normal ageing in the column/pre-column. If the pressure is too high, the column can be flushed or the precolumn replaced. As the pressure only increases slowly, this problem is often not immediately apparent. If there is a sudden increase in pressure, it is advisable to check whether maintenance work or something else has been carried out on the system beforehand. This may give an indication of the cause.

 

ObservationPossible causeSolution/Prevention
Pressure builds up graduallyAgeing of the column/column

1. flush the column

2. if flushing does not show any improvement, replace the column/pre-column

Pressure is suddenly too highClogged frit

1. flush the column in reverse (please observe the manufacturer's information beforehand)

2. use inline filter

3. if possible: replace the inlet frit of the column (please observe the manufacturer's information beforehand)

Pressure is suddenly too highClogged capillary1. systematically check capillaries
Pressure is suddenly too highClogged inline filter1. replace inline filter
Pressure is suddenly too highFlow rate too high1. check the pump settings
Pressure is suddenly too highTemperature too low, resulting in higher viscosity1. check the temperature of the column oven
Pressure is suddenly too highBuffer/salt has failed

1. check solubility of the additives used in all solvent compositions before using a new method

2. rinse the column

Pressure is suddenly too highWrong column (smaller inner diameter, longer or smaller particle size)1. check the column used
Pressure is suddenly too highIncorrect capillaries used, can happen during maintenance/replacement of capillaries

1. check capillaries used

2. document all maintenance/replacement work

Pressure is suddenly too highStainless steel fittings that are too tight can press in the capillary and create a bottleneck

1. tighten fittings according to manufacturer's instructions

2. replace capillaries/connections

Pressure is suddenly too highIncorrect solvent composition

1. check the mixed mobile phase

2. with gradient valve, check valves for functionality


Pressure too low

If the pressure is too low, this is almost always due to a leak in the system. If the pressure is too low, you should therefore always check for leaks first. In addition to leaks, flush valves can also be the problem if they have not been closed after flushing.

 

ObservationPossible causeSolution/Prevention
Pressure too lowLeak in the system

1. search for leaks and repair connection

1. no obvious leak: check all connections

Pressure too lowFlush valve not closed1. close flush valve
Pressure too lowPump non-return valve defective

1. clean non-return valve (observe manufacturer's instructions)

2. replace non-return valve

Pressure too lowFlow rate too low1. check pump settings
Pressure too lowTemperature too high, resulting in lower viscosity1. check column oven
Pressure too lowNo solvent in the supply so that the pump cannot deliver anything

1. check solvent supply

2. flush air bubbles out of the pump if necessary

Pressure too lowWrong solvent with lower viscosity used1. check solvent/mobile phase
Pressure too lowIncorrect solvent composition

1. check mixed mobile phase

2. with gradient valve, check valves for functionality


Baseline problems

Problems with the base line can be characterised in different ways. On the one hand there are randomly occurring problems and on the other hand cyclically occurring problems. Random problems are usually caused by impurities or air bubbles. This can even manifest itself as an additional peak in the form of spikes. Problems that occur cyclically or periodically indicate a problem with the detector, such as baseline drift or a temperature fluctuation.

 

ObservationPossible causeSolution/Prevention
Baseline driftDetector not sufficiently stabilised1. flush detector with mobile phase until stable baseline is achieved (especially with gradient methods)
Baseline driftGradient method

1. some detectors cannot be operated with a gradient (e.g. RI detector)

2. a certain drift can be normal with gradients

Baseline driftContamination in the detector cell1. flush the detector cell (follow the manufacturer's instructions)
Baseline driftTemperature fluctuations during a run1. check temperature settings of the detector cell/column oven
Spikes in the baselineAir bubbles in the system1. flush the system
Spikes in the baselineImpurities in the column/mobile phase

1. rinse column

2. prepare new mobile phase

Spikes in the baselineEnergy of the detector lamp no longer sufficient1. replace detector lamp
Strong noise in the baselineSensitivity of the detector set too high1. check the detector settings
Other baseline problemsProblem with the electronics of the detector1. contact manufacturer

Downloads

Contact Us

Fields marked with * must be filled in

The right column for you - we will be happy to support you individually

Competent consultants are always at your side. Write a message to our consultants, we will get back to you and give you individual support.

You will find:

Application support Application support
Availability and delivery times Availability and delivery times
Price information with individual conditions Price information with individual conditions
Certificates of analysis, MSDS, data sheets Certificates of analysis, MSDS, data sheets