For applications in the field of preparative chromatography, MZ-Analysentechnik GmbH supplies ready-packed HPLC columns with almost the entire range of chromatographic materials in dimensions from 8.0 to 50.0 mm ID (semi-prep - prep). In the manufacture of our in-house preparative columns, we use a specially optimised packing procedure for each packing material and each pre-packed preparative column is then checked to ensure that it meets our strict quality requirements.

Products

Sort By sort-descending

Items 21-30 of 2301

Page
per page
MZ0843-150100
MZ0843-250080
MZ0843-150080
MZ0843-100080
MZ0830-250400
MZ0830-250300
MZ0830-250200
MZ0830-250100
MZ0830-150100
MZ0830-250080
Sort By sort-descending

Items 21-30 of 2301

Page
per page

Technical Data

How do I choose the right preparative column?

In principle, the stationary phase should already be defined. Normally, experiments are carried out with analytical columns and these methods are then scaled up to the preparative scale. It is important to ensure that the particle size ideally remains the same. If a larger particle size is used for the preparative column, it is advisable to maintain the ratio of particle size to column length. In addition, the mobile phase and the solvent of the sample should also be identical.

 

Therefore, when selecting the correct preparative column, the question of which dimension is required should be asked. As maximum loadability is desired in preparative chromatography, it is advisable to first optimise the analytical method and achieve the highest possible resolution. Once the highest resolution has been achieved, the loadability can be analysed:

 

  1. A sample with a high concentration is prepared and the injection volume is increased.
  2. Several samples with different concentrations are prepared and the injection volume remains constant

 

This should maximise the sample quantity so that the peaks to be separated can still be separated with sufficient purity. This maximum loadability varies from column to column and is highly dependent on the individual separation problem and resolution. This loadability must therefore be determined empirically for each separation. The resulting poor peak shapes or overloading of the detector can be ignored during such preparative preparation.


The maximum loading capacity of the analytical column has been determined - what happens now?

Once the maximum load (defined amount of analyte in a defined injection volume) has been determined, you should consider how much of the analyte should be separated in one run. This injection volume can then be used to calculate the appropriate internal diameter using the following formula. If the particle size of the analytical and preparative column is identical and therefore the length remains the same, the length term becomes one and can be neglected.

VInj,Prep = VInj,Analy - ( LPrep - dc Prep 2 LAnaly - dc analyte 2 )

Further adjustments

With the appropriate inner diameter of the preparative column, the remaining parameters such as the flow velocity and possibly the gradient time can also be calculated.

Adjusting the flow velocity

FPrep = FAnaly - ( dpAnaly - dc Prep 2 dpPrep - dc Analy 2 )

Adjustment of the gradient

Different systems are often used for an analytical and preparative method. This also results in different dwell and dead times (dwell volume and dead volume). To compensate for this, short isocratic steps can be incorporated at the start of the gradient. The following equation applies to gradients that are transferred from one system to another:

t D,Analy + t I,Analy t c,Analy = t D,Prep + t I,Prep t c,Prep

tD,Analy Dwell time for the analytical system

tI,Analy Time of the isocratic step at the beginning of the gradient for the analysed system system

tc,Analy Dead time for the analytical system

tD,Prep Dwell time for the preparative system

tI,Prep Time of the isocratic step at the beginning of the gradient for the prep. system system

tc,Prep Dead time for the preparative system


Compensation of different dwell volumes of both systems

In order to compensate for different residence volumes between an analytical and preparative system, short isocratic steps can be incorporated before the actual gradient. The time of such a step can be calculated using the following formula, whereby the residence volumes as well as the column dead volume must be known for both systems.

t I,Prep = ( t I,Analy - F Analy ν C,Analy + ν D,Analy ν C,Analy - ν D,Prep ν C,Prep ) - ν C,Prep F Prep

tI,Prep Time of the isocratic step at the beginning of the gradient for the prep. system. system

tI,Analy Time of the isocratic step at the beginning of the gradient for the analy. system system

FPrep Flow rate for the preparative system

FAnaly Flow rate for the analytical system

vC,Prep column dead volume for the preparative system

vD,Prep Dwell volume for the preparative system

vC,Analy column dead volume for the analytical system

vD,Analy Residual volume for the analytical system


With all these calculations, you can scale up your analytical method to a preparative method. In this process you can also use our HPLC method transfer tool can also support you in this process. We are also happy to help you with the calculations. Get in touch with us! At MZ-Analysentechnik GmbH, you can also obtain any material as a preparative column to ensure seamless scaling from analytical to preparative methods.

Downloads

Contact Us

Fields marked with * must be filled in

The right column for you - we will be happy to support you individually

Competent consultants are always at your side. Write a message to our consultants, we will get back to you and give you individual support.

You will find:

Application support Application support
Availability and delivery times Availability and delivery times
Price information with individual conditions Price information with individual conditions
Certificates of analysis, MSDS, data sheets Certificates of analysis, MSDS, data sheets